Validation of dual energy X-ray absorptiometry measures of abdominal fat by comparison with magnetic resonance imaging in an Indian population.

Amy ETaylor; Hannah Kuper ORCID logo; Ravi DVarma; Jonathan CWells; Jimmy DBell; KV Radhakrishna; BharatiKulkarni; Sanjay Kinra ORCID logo; Nicholas JTimpson; ShahEbrahim; +2 more... George DaveySmith; YoavBen-Shlomo; (2012) Validation of dual energy X-ray absorptiometry measures of abdominal fat by comparison with magnetic resonance imaging in an Indian population. PloS one, 7 (12). e51042-. ISSN 1932-6203 DOI: 10.1371/journal.pone.0051042
Copy

OBJECTIVE: Abdominal adiposity is an important risk factor for diabetes and cardiovascular disease in Indians. Dual energy X-ray absorptiometry (DXA) can be used to determine abdominal fat depots, being more accessible and less costly than gold standard measures such as magnetic resonance imaging (MRI). DXA has not been fully validated for use in South Asians. Here, we determined the accuracy of DXA for measurement of abdominal fat in an Indian population by comparison with MRI. DESIGN: 146 males and females (age range 18-74, BMI range 15-46 kg/m(2)) from Hyderabad, India underwent whole body DXA scans on a Hologic Discovery A scanner, from which fat mass in two abdominal regions was calculated, from the L1 to L4 vertebrae (L1L4) and from the L2 to L4 vertebrae (L2L4). Abdominal MRI scans (axial T1-weighted spin echo images) were taken, from which adipose tissue volumes were calculated for the same regions. RESULTS: Intra-class correlation coefficients between DXA and MRI measures of abdominal fat were high (0.98 for both regions). Although at the level of the individual, differences between DXA and MRI could be large (95% of DXA measures were between 0.8 and 1.4 times MRI measures), at the sample level, DXA only slightly overestimated MRI measures of abdominal fat mass (mean difference in L1L4 region: 2% (95% CI:0%, 5%), mean difference in L2L4 region:4% (95% CI: 1%, 7%)). There was evidence of a proportional bias in the association between DXA and MRI (correlation between difference and mean -0.3), with overestimation by DXA greater in individuals with less abdominal fat (mean bias in leaner half of sample was 6% for L1L4 (95%CI: 2, 11%) and 7% for L2L4 (95% CI:3,12%). CONCLUSIONS: DXA measures of abdominal fat are suitable for use in Indian populations and provide a good indication of abdominal adiposity at the population level.



picture_as_pdf
pone.0051042.pdf
subject
Published Version
Available under Creative Commons: 3.0

View Download

Explore Further

Read more research from the creator(s):

Find work associated with the faculties and division(s):

Find work associated with the research centre(s):

Find work from this publication:

Find other related resources: