Diagnostic Accuracy of the Biosynex CryptoPS Cryptococcal Antigen Semiquantitative Lateral Flow Assay in Patients with Advanced HIV Disease.

Mark WTenforde; TimothéeBoyer-Chammard; CharlesMuthoga; LeabanengTawe; ThandiMilton; IkanyengRulaganyang; KwanaLechiile; IvyRukasha; Tshepo BLeeme; Nelesh P Govender ORCID logo; +6 more... JuliaNgidi; MadisaMine; Síle FMolloy; Thomas SHarrison; OlivierLortholary; Joseph N Jarvis ORCID logo; (2021) Diagnostic Accuracy of the Biosynex CryptoPS Cryptococcal Antigen Semiquantitative Lateral Flow Assay in Patients with Advanced HIV Disease. JOURNAL OF CLINICAL MICROBIOLOGY, 59 (1). ISSN 0095-1137 DOI: 10.1128/JCM.02307-20
Copy

High cryptococcal antigen (CrAg) titers in blood are associated with subclinical meningitis and mortality in CrAg-positive individuals with advanced HIV disease (AHD). We evaluated a novel semiquantitative lateral flow assay (LFA), CryptoPS, that may be able to identify individuals with high CrAg titers in a cohort of AHD patients undergoing CrAg screening. In a prospective cohort of patients with AHD (CD4 cell count, ≤200/μl) receiving CD4 count testing, whole blood was tested for CrAg by CryptoPS and the IMMY LFA; the two assays were conducted by two different operators, each blind to the results of the other assay. The sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of CryptoPS were assessed against the IMMY LFA as a reference. CryptoPS low-titer (T1 band) and high-titer (T2 band) results were compared with IMMY LFA titers obtained through serial dilution. A total of 916 specimens were tested. The sensitivity of the CryptoPS assay was 61.0% (25/41) (95% confidence interval [95% CI], 44.5 to 75.8%), its specificity was 96.6% (845/875) (95% CI, 95.1 to 97.7%), its PPV was 45.5% (95% CI, 32.0 to 59.4%), and its NPV was 98.1% (95% CI, 97.0 to 98.9%). All (16/16) CryptoPS false-negative results were obtained for samples with IMMY titers of ≤1:160. Of 29 patients (30 specimens) who tested positive by CryptoPS but negative by the IMMY LFA, none developed cryptococcal meningitis over 3 months of follow-up without fluconazole. Median CrAg titers were 1:20 (interquartile range [IQR], 0 to 1:160) in CryptoPS T1-positive samples and 1:2,560 (IQR, 1:1,280 to 1:10,240) in T2-positive samples. We conclude that the diagnostic accuracy of the CryptoPS assay was suboptimal in the context of CrAg screening, with poor sensitivity at low CrAg titers. However, the CryptoPS assay reliably detected individuals with high titers, which are associated with poor outcomes.



picture_as_pdf
JCM.02307-20.pdf
subject
Published Version
Available under Creative Commons: 3.0

View Download

Explore Further

Read more research from the creator(s):

Find work associated with the faculties and division(s):

Find work from this publication: